Technical Specification
1. Flow rate:1ml/h~500ml/h, 1ml/h increment.
2. Target infused volume / Tar. volume: 0~9999ml, (0~999.9ml, 0.1ml increment; 1000~9999ml, 1ml increment).
3. Infused volume: 0~9999ml, (0~999.9ml, 0.1ml increment; 1000~9999ml, 1ml increment).
4. The flow precision is ± 10% according to the clinical setting.
5. Purge and reverse rate: 500ml/h.
6. Fuse specification: F1AL /250V, 2 pieces per pump.
7. Power supply: AC 100V~240V, 50/60Hz, ≤25VA,
8. Internal Lithium-ion polymer battery: DC 11.1V, ≥1600mAh. The battery backup is approx. 5 hours when using a fully charged battery and running at 25ml/hr.
9. Operating condition:
Temperature 5~40oC,
Relative humidity: 20~90%.
Atmospheric pressure: 860hPa~1060hPa.
10. Safe grade: Class I, type CF
11. Size: 130 mm (W) × 150mm (H) × 100mm (D)
12. Net weight: approx. 2kg